K0430 Issue No. 4 24
Select Output
To display an output channel, select ADVANCED mode from the task menu. The display
shows the list of output selections and, if available, the UNITS soft box (F2).
The following procedure shows the method of output channel selection from two pages
of options. The second page can be obtained directly from the first by pressing the right
cursor key:
Note: Left/right arrow keys function as page up/down keys.
To change the output units (pressure channels only), select the channel with the cursor
keys and press F2 before pressing ENTER.
Electrical Outputs (Loop Power)
All the electrical outputs, the output loop must be powered from an external supply
(current sinking).
External connections to the front panel of the instrument are shown below.
Current Sinking
A Load B Barrier C (External supply 30 V dc maximum) D Safe area
+
-
B
C
D
B
55mA
Vin
mA in
mA Sink
max 30V
CAT II
A
max
Advanced Task Select Output